The reported event was confirmed manufacturing related.The product had caused the reported failure.Sample was evaluated and observed there was a pinhole on the bottom part of sac collar that allowed water to leak out.Pinhole was examined under microscope and noted to be round and smooth.A review of the manufacturing process indicates that the process can produce of this type of defect.Therefore, this failure mode confirmed manufacturing related.A potential root cause for this failure could be due to bubble during dipping process.The labeling is found to be adequate.No non conformances were identified in any of the dhrs.Therefore, no additional action required at this time.Correction: d,h
udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|