A lead extraction procedure commenced to remove a right atrial (ra) and right ventricular (rv) lead due to bacteremia.During preparation, the rv lead was removed successfully with manual traction.A spectranetics lld e lead locking device (lld e) was inserted into the ra lead to provide traction.Beginning with a spectranetics 11f tightrail sub-c rotating dilator sheath, it was noted that the lead was heavily calcified and resistance was encountered.Advancement was made just before the superior vena cava (svc), and upsized to a spectranetics 13f tightrail rotating dilator sheath.Resistance was observed on the entire length of the lead, but the ra lead was freed and extracted.However, the patient's blood pressure dropped, and rescue efforts began, including rescue balloon, cpr, blood product, medications, and sternotomy.An ra perforation was discovered and repaired, and the patient survived the procedure.This report captures the lld e providing traction on the ra lead when the perforation occurred, requiring intervention.There was no alleged malfunction of any spectranetics devices in use during the procedure.
|
|
A3b) patient gender - not available from facility.A4) patient weight - not available from facility.B2) due to multiple failed submissions for entering 2 outcomes, adding here the additional outcomes attributed to adverse event: life threatening.B7) other relevant history - not available from facility.H3) the lld e was discarded, thus no product investigation was performed.H6) per ifu, perforation is listed as a potential adverse event with use of the lld device.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
|