A patient implanted with a talisman implant kit on (b)(6) 2025 reported during a follow up appointment ((b)(6) 2026) that the incision was not fully healed.The stimrouter implantable lead and lead introducer kit were utilized during the talisman implantation.Decision by the patient's physician was to leave the dissolvable stitch for wound closure in place.Redness and swelling increased.While in ny, patient visited emergency room for worsening local symptoms.The patient was prescribed clindamycin and recommended to return to (b)(6) for follow up care as well as a potential explant.The stimrouter lead was explanted without issue (b)(6) 2026, however the talismann (implantable pulse generator kit) ipg became detached during the explant process.The patient was referred to the implanting physician for surgical excision of the talismann ipg in the operating room.On (b)(6) 2026 the device was successfully excised.The implant of the talismann model tm-1000 pns device utilized the stimrouter model st2-1000 stimrouter implantable lead and lead introducer kit, sterile.This report is for the talismann model tm-1000 pns device.A seperate report is submitted for the st2-1000 stimrouter implantable lead and lead introducer kit, sterile.
|
|
More than a week following a talismann implant on (b)(6) 2026 the incision had not fully healed and the non-absorbable stitch (presumably used to close the skin) was "left" in place, which is standard for an absorbable stitch.The patient increasingly showed signs of an infection including redness leading to presentation at an emergency room.An explant was recommended but during this the ipg (talismann) detached and had to be separately removed.Photos of the affected site show evidence of an infection with induration, redness, and pus.The wound appears not to have healed, and the stimrouter lead can be clearly seen in the wound.Based on the implant protocol this is likely the proximal aspect of the lead.It is assumed that these are photos taken prior to the initial explanation attempt.Because serious injury occurred and the talismann device/procedure did contribute to the development of this serious injury (this is a known side effect, there is a temporal relationship, the event involves a part of the body where the device was applied, and harm occurred).In this case, there are two issues that occurred.The first is infection.Infection can occur due to a number of factors including patient factors (i.E., history of mrsa/staph infections suggesting that the patient may have an underlying susceptibility and also patient behavior post-implant) and clinician factors (i.E., surgical technique).There is unlikely to be an inherent issue with the device that promoted infection.The second issue that occurred is incomplete explant with separation of the lead from the talismann ipg.In this case, because the lead separated from the talismann ipg, a second surgery was necessary.This is an event that would lead to injury if the malfunction were to recur.There are different factors that could have contributed to this particular event - clinician factors (i.E., not following the clinician procedure manual including proper deployment of the silicone sleeve and use of the 2-0 suture) and device factors (failure of the silicone sleeve or suture to withstand tension when removing the device).It is notable that the procedure manual (in terms of explant) states that if explantation occurs beyond 14 days after implant, the technique is at the discretion of the explanting physician although the recommended steps are "still applicable." the technique specifies cut-down to expose the silicone sleeve of the ipg.As the ipg detached during the initial explant surgery.It can reasonably be inferred that if the removal of talismann required a second surgery then the cut-down technique suggested in the manual may not have been used.Therefore, a second surgery was necessary to remove the talismann ipg.The investigation of the involved talismann implant kit production lot records for sterility and production issues potentially associated with the stimrouter model st2-1000 implantable lead separation from the talismann model tm-1000 pns device is ongoing.
|