It was reported that a patient underwent a cardiac ablation and the patient experienced cardiac tamponade that required drainage.Cavotricuspid isthmus (cti) ablation was performed using the qdot micro catheter and vizigo short.Brockenbrough (bb) was performed using the soundstar eco catheter.Pulmonary vein isolation (pvi) was performed using the varipulse catheter, and post-mapping of the left atrium (la) was performed using octaray.After completion of these procedures, the physician observed a decrease in blood pressure, confirmed using echocardiography, which revealed pericardial effusion.The patient's blood pressure recovered and stabilized after drainage, and the patient left the room.It was stated that the drained blood was venous blood, so the issue might have been related to ablation near the inferior vena cava (ivc), during cti ablation, or to coronary sinus (cs) catheter insertion.Visitag coloring setting was tag index.Visitag additional filter used was fot.Force visualization features were used were cf, realtime graph, vector, visitag.Parameters for stability were: range: 3mm, time: 3sec, fot: 25%5g, tag size: 2mm.Procedural activated clotting time was 350 seconds over.One catheter was used for each exchange.Transseptal puncture was performed, one site.There was no evidence of steam pop.Flow setting on the irrigated catheter was pre: 2 sec, post: 5 sec.There were no abnormalities observed prior to, nor during use of the product.
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
|