The instructions for use for the ijs-e system states the following: "the locking screws of the construct and the axis pin must be installed and fully tightened to ensure that the construct will maintain the positioning and angles established intraoperatively.If the locking screws or the axis pin are not attached and/or fully tightened, the construct may loosen, shift and/or become disassembled subcutaneously." "the ijs-e construct connecting rod splines must be fully seated prior to tightening.Improper positioning may result in loosening of the construct." "improper placement, positioning, alignment or fixation of the ijs-e construct may result in unusual stress conditions which may lead to subsequent reduction in the service life of the components, construct failure, postoperative complications or ineffective treatment." "for safe effective use of the implant, the surgeon must be thoroughly familiar with the surgical technique, implant, and associated instruments.Improper insertion of the device during implantation may also increase the possibility of loosening, migration and failure of the device or the treatment." review of the images from the case found that the axis pin was not positioned properly in accordance with the surgical technique.Additionally, examination of the returned device revealed damage to the connecting rod splines consistent with not fully seating the splines prior to tightening.
|