• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION 303, INC. Alaris System; Pump, Infusion

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION 303, INC. Alaris System; Pump, Infusion Back to Search Results
Model Number 8015
Medical Device Problem Code Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/16/2026
Type of Reportable Event Malfunction
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Alaris System
Common Device Name
Pump, Infusion
Manufacturer (Section D)
CAREFUSION 303, INC.
3750 torrey view ct
san diego CA 92130
Manufacturer (Section G)
CareFusion 303, INC.
10020 pacific mesa blvd
san diego CA 92121 -438
Manufacturer Contact
brett wilko
3750 torrey view ct
san diego, CA 92130
+1(858)617-2000
MDR Report Key50096956
Report Number2016493-2026-50683
Device Sequence Number21400369
Product Code FRN
UDI-Device Identifier10885403801532
UDI-Public(01)10885403801532
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K072105
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number8015
Device Catalogue Number8015 Alaris PCU 1.5 Module
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/15/2026
Initial Report FDA Received Date07/31/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/06/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-