Medtronic Mexico S. De R.L. De CV VitalFlow Set; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIO
|
Back to Search Results |
|
| Model Number |
58112 |
| Medical Device Problem Codes |
Obstruction of Flow (2423); Communication or Transmission Problem (2896); Noise, Audible (3273)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/19/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Medtronic received information that during use of a vital flow set, it was reported that at initiation, the circuit was presented with e49 error code and an intermittent/unreliable pressure and saturation readings.After 4 days of support without patient support issues, the circuit suddenly went to zero flow with a loud rattling sounds emanated from the pump head.Pump was transitioned to hand crank and back to motor, and was able to achieve flow with loud rattle persisting.Pump went to zero flow once more and then a decision was made to perform an urgent circuit exchange.The device was replaced to complete the procedure.There was no adverse patient effect associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|