MEDTRONIC PUERTO RICO OPERATIONS CO. Evera MRI XT DR SureScan; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number DDMB1D1
Medical Device Problem Codes Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574); Device Sensing Problem (2917)
Health Effect - Clinical Code Shock from Patient Lead(s) (3162)
Date of Event 07/13/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Continuation of d10: 5076-52 lead implanted (b)(6) 2012.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the right atrial (ra) lead exhibited undersensing during atrial tachycardia/atrial fibrillation (at/af) events resulting in unnecessary pacing at times.Intermittent undersensing also resulted in misclassification of a detected ventricular tachycardia (vt) episode that was treated with anti-tachycardia pacing (atp) therapy.The ra lead remains in use. it was also reported that the patient received atp and high voltage therapy from the cardiac resynchronization therapy defibrillator (crt-d) for possible atrial tachycardia/atrial fibrillation (at/af) with fast ventricular response that the device detected as vt.There were also monitored episodes of possible at/af with fast ventricular response that were detected as non-sustained ventricular tachycardia.Functional undersensing in blanking/refractory on the ventricular channel also resulted in overpacing.The crt-d remains in use. no further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Evera MRI XT DR SureScan
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112--439
+1(763)526-9277
MDR Report Key50098384
Report Number3004209178-2026-12853
Device Sequence Number15906680
Product Code LWS
UDI-Device Identifier00643169837713
UDI-Public(01)00643169837713(17)200728
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P980016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/31/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date07/28/2020
Device Model NumberDDMB1D1
Device Catalogue NumberDDMB1D1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/14/2026
Initial Report FDA Received Date07/31/2026
Date Device Manufactured02/06/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
694765 LEAD
Outcome Attributed to Adverse Event Life Threatening;
Patient Age76 YR
Patient SexFemale
-
-