| Model Number |
E200-31M |
| Medical Device Problem Codes |
Perivalvular Leak (1457); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported on (b)(6) 2026, a 31mm epic plus mitral valve was chosen for a procedure.After ligation, the cuff and shield sections were generally sewn together with uniform stitching; however, a portion was found to have non-uniform spacing, with a gap of approximately 1.5 stitch intervals.There was para leak was noted.The valve remains implanted and has not been explanted.The patient was reported stable.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of non-uniform stitching and paravalvular leak was reported.A returned device inspection, to rule out any device-related causes, could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all specifications.Based on the available information, the cause of the reported non-uniform stitching and subsequent paravalvular leak could not be conclusively determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
|
| |
|