A male patient underwent aquablation therapy for symptomatic benign prostatic hyperplasia (bph).Procept biorobotics (procept) became aware that post-aquablation therapy, the patient experienced postoperative bleeding.The patient received a blood transfusion of one unit.After transfusion, no surgical intervention happened, patient was admitted and discharged the next day.No malfunction of the hydros robotic system was reported.
|
|
The hydros robotic system is a reusable device; therefore, it is currently in the possession of the user facility.The investigation of this event consisted of a review of the device history record (dhr), and instructions for use (ifu).A review of the device history record (dhr) for hy1000/serial number 25c01730 was conducted, which confirmed that there were no non conformances, failures, discrepancies, or missed steps during the manufacturing process that could be related to the reported event.The review indicated that the device met all design and manufacturing specifications when released for distribution.A review of the log files for this procedure could not be conducted as they were not provided.Multiple attempts were made to obtain the log files without success.The hydros robotic system's user manual (um), um0401-00-01 rev.C was reviewed.4.2.3 warnings: procedure
as with any surgical urologic procedure, potential perioperative risks of the aquablation procedure include the following, some of which may lead to serious outcomes and may require intervention:
o bleeding or blood in urine
section 9: post-op sterile, hemostasis states the following:
prior to introducing the resectoscope to start hemostasis, reduce trus compression and ensure there is no steep angling of the trus probe.Note: depending on patient anatomy, consider that trus probe may be removed entirely, if needed to manipulate the resectoscope adequately for hemostasis.Warning: avoid applying excessive compression to the prostate.Excessive compression can contribute to serious injury to the patient.Prepare the resectoscope and toomey syringe/clot evacuator for hemostasis.Introduce resectoscope sheath.Evacuate clots with toomey or ellik evacuator.Caution: position the electrosurgical generator apart from the hydros robotic system, as permitted by the operating room layout.Use a loop to systematically work around the bladder neck and remove the ¿fluffy¿ tissue (tissue devitalized by aquablation therapy) to view potential bleeding vessels underneath.Perform focal cautery at bleeders.Turn off irrigation and inspect for bleeders under normal blood pressure.Note: alternate hemostasis methods may include:
- balloon catheter in bladder with bladder neck traction.- balloon catheter in bladder with bladder neck traction.Fill the bladder with sterile saline and maintain for 30-60 minutes before starting cbi.Under spinal anesthesia.- balloon catheter in bladder, no traction.- balloon catheter in prostatic fossa.- inflate balloon with 5cc saline in the bladder.- retract balloon under trus guidance into prostatic fossa.- inflate balloon to 30-50% of the initial prostate volume.- apply mild traction to hold the balloon catheter in place.Once hemostasis is achieved, remove trus probe and inspect probe and anus for blood and investigate, as needed.Introduce 22-24f hematuria catheter, inflate balloon, and apply standard to light traction as needed.Run cbi per facility's turp protocol.Monitor irrigation bags to maintain fluid supply.Monitor effluent color.Warning: unaddressed vascular bleeding can lead to serious patient harm.The hydros robotic system is a reusable device; therefore, it is still currently in possession of the user facility.The hydros robotic system's user manual lists bleeding as potential risks of aquablation therapy and provides adequate instructions on how to achieve appropriate hemostasis.It was reported that the patient was bleeding and received a blood transfusion of 1 unit.After transfusion, no surgical intervention happened, patient was admitted and discharged next day.No malfunction of the hydros robotic system was reported.Based on the review of the information provided, dhr, and um, the event is considered not to be device related.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
|