• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PROCEPT BioRobotics Corporation AQUABEAM Robotic System; Fluid Jet Removal System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PROCEPT BioRobotics Corporation AQUABEAM Robotic System; Fluid Jet Removal System Back to Search Results
Catalog Number AB2000
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Perforation (2001)
Date of Event 05/08/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The aquabeam robotic system is a reusable device; therefore, it is currently in the possession of the user facility.The investigation of this event consisted of a review of the device history record (dhr), treatment log files and instructions for use (ifu).A review of the device history record (dhr) for ab2000/serial number: (b)(6) was conducted, which confirmed that there were no non conformances, failures, discrepancies, or missed steps during the manufacturing process that could be related to the reported event.The review indicated that the device met all design and manufacturing specifications when released for distribution.The aquabeam robotic system's treatment log file was reviewed, which confirmed no malfunctions during the aquablation procedure and indicated that the system functioned as designed.The aquabeam robotic system's instruction for use (ifu), sj-ifu0104-00, rev.B, was reviewed.4.3.Warnings: procedure: as with any surgical urologic procedure, potential perioperative risks of the aquablation procedure include: ¿ bladder or prostate capsule perforation.The aquabeam robotic system is a reusable device; therefore, it is still in the user facility's possession.It was reported that the patient sustained a slight bladder perforation.The patient had a catheter in place before aquablation therapy and a pre-existing diverticulum.It was also noted that the patient had lots of calcifications that potentially distorted the ultrasound image, and cautery took an extended time due to an erection and leftover apical tissue.The patient was then sent home, but later fell ill, and a ct scan revealed leakage in the bladder.The treating surgeon is unsure of the origin of the leakage and has no other information.No additional surgery or medical intervention was required.The patient is being managed by catheter and cystogram and has been discharged.No additional surgery or medical intervention was required.The aquabeam robotic system's ifu lists bladder perforation as a potential risk of aquablation therapy.Based on the information received and a review of the treatment log files, dhr, and ifu, the event is considered not device-related.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Event or Problem Description
A male patient underwent aquablation therapy to treat symptomatic benign prostatic hyperplasia (bph) on (b)(6) 2026.Procept biorobotics became aware on (b)(6) 2026 that the patient sustained a slight bladder perforation.The patient had a catheter in place before aquablation therapy and a pre-existing diverticulum.It was also noted that the patient had lots of calcifications that potentially distorted the ultrasound image, and cautery took an extended time due to an erection and leftover apical tissue.The patient was then sent home, but later fell ill, and a ct scan revealed leakage in the bladder.The treating surgeon is unsure of the origin of the leakage and has no other information.No additional surgery or medical intervention was required.No malfunction of the aquabeam robotic system was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AQUABEAM Robotic System
Common Device Name
Fluid Jet Removal System
Manufacturer (Section D)
PROCEPT BioRobotics Corporation
150 baytech dr
san jose CA 95134
Manufacturer (Section G)
PROCEPT BioRobotics Corporation
150 baytech dr
san jose CA 95134
Manufacturer Contact
doria esquivel
150 baytech dr
san jose, CA 95134
+1(650)232-7291
MDR Report Key50098621
Report Number3012977056-2026-00155
Device Sequence Number10271338
Product Code PZP
UDI-Device Identifier00850055427020
UDI-Public0100850055427020112405312124C01594
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
DEN170024
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/31/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberAB2000
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 07/10/2026
Initial Report FDA Received Date07/31/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/31/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-