| Model Number |
CDS0801-XT |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Mitral Valve Insufficiency/ Regurgitation (4451); Unspecified Tissue Injury (4559)
|
| Date of Event |
07/08/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026, a patient presented with grade 4 degenerative mitral regurgitation (mr), flail, prolapsed posterior and anterior leaflet for a mitraclip procedure.During the procedure, first and second mitraclips was implanted and physician decided to implant third mitraclip to further reduce the mr.When attempting to grasp the leaflet with the third clip, it was noted that there was possible perforation of the posterior leaflet.Third clip was implanted.The final mr was grade 4.There were no adverse patient effects or clinically significant delays reported.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6)2026, a patient presented with grade 4 degenerative mitral regurgitation (mr), flail, prolapsed posterior and anterior leaflet for a mitraclip procedure.During the procedure, first and second mitraclips was implanted and physician decided to implant third mitraclip to further reduce the mr.When attempting to grasp the leaflet with the third clip, it was noted that there was possible perforation of the posterior leaflet.Third clip was implanted.The final mr was grade 4.There were no adverse patient effects or clinically significant delays reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Based on available information, the cause of the reported tissue injury was unable to be determined.The reported unchanged mr was likely a cascading effect of the reported tissue injury.The reported patient effects of tissue injury and mitral regurgitation, as listed in the mitraclip system instructions for use, are known possible complications associated with mitraclip procedures.The reported serious injury/ illness/ impairment was a result of case-specific circumstances.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|