Note: related naviflex long wire pusher complaints are reported under records (b)(4)and (b)(4).It was reported to boston scientific corporation that a naviflex long wire pusher was to be used during an unknown procedure performed on (b)(6)l 2026.During preparation for the procedure, a labeling discrepancy was identified.The package was labeled as a long wire pusher; however, the device contained within the package was a short wire pusher.The product previously used by the facility was a long-wire type, in which the guidewire passes directly through the inlet and outlet.The product received on this occasion was confirmed to be an rx type, in which the guidewire exits through a side port located on the outer sheath.There were no patient complications reported as a result of this event.
|
|
Block h6:
imdrf device code a2101 captures the reportable event of device labelling issue.Block h11:
investigation results:
based on the information available, boston scientific confirmed the reported event of a device labeling issue.Product analysis determined that the pusher included an rx port.However, based on the upn (m00535030), the product specifications for this device do not include an rx port.Device history record review:
it was confirmed this device met manufacturing specifications prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Device technical analysis:
the returned advanix pancreatic stent was analyzed.The pusher was observed to contain an rx port.However, according to the specifications associated with upn m00535030, this device configuration should not include an rx port.No other problems with the device were noted.Labeling review:
a labeling review was performed.The ifu contains detailed device information and instructions for the device use.There is no evidence the device was improperly used per the ifu; however, although the label information corresponds to a long wire pusher device, the device contained inside does not match the configuration or specifications of a long wire pusher.Investigation conclusion:
based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of manufacturing deficiency.Boston scientific has been notified that the advanix pancreatic stent naviflex long wire pusher (upn: m00535030) contained a pusher component that was inconsistent with the product labeling.A ship hold has been placed on the lot associated with this complaint device.Boston scientific has also initiated a non-conforming events prevention (ncep) investigation to further evaluate the reported event, and determine appropriate actions.
|