| Model Number |
TCDS0801-XT |
| Medical Device Problem Codes |
Improper or Incorrect Procedure or Method (2017); Incomplete Coaptation (2507)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/08/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026, a patient presented with grade 4 functional tricuspid regurgitation (tr) and rotated heart for a triclip procedure.During the procedure, first triclip was implanted and second triclip was implanted on the medial side of posterior and septal leaflet.It was determined that there was a single leaflet device attachment (slda) and the clip was no longer attached to the septal leaflet.Third triclip was implanted to stabilize the slda.Imaging was difficult.The final tr was grade 2.There were no adverse patient effects or clinically significant delays reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot specific product quality issue.Based on the information reviewed, the reported single leaflet device attachment (slda) appears to be related to a combination of patient anatomy and procedural conditions.The reported visualization difficulty appears to be related to procedural conditions.The cause of the reported ifu deviation appears to be related to procedural conditions.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|
| |
|