Axonics Modulation Technologies, Inc. Axonics; Stimulator, Electrical, Implantable, For Incontinence Back to Search Results
Model Number 1101
Medical Device Problem Codes Premature Discharge of Battery (1057); Charging Problem (2892)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/07/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient with the neurostimulator underwent an explant due to the battery failing to charge.The patient no longer wanted to charge their battery therefore they opted for a replacement.The device was replaced with a different model device.The patient states that they've had 2 years with issues charging the device and that the battery was dead for a while before they decided to change the model.Only the neurostimulator was revised.
 
Event or Problem Description
It was reported that the patient with the neurostimulator underwent an explant due to the battery failing to charge.The patient no longer wanted to charge their battery therefore they opted for a replacement.The device was replaced with a different model device.The patient states that they've had 2 years with issues charging the device and that the battery was dead for a while before they decided to change the model.Only the neurostimulator was revised.
 
Additional Manufacturer Narrative
Block d2b: additional pro code: qon block g4: additional premarket / 510 (k): p190006 the device was not returned for analysis; therefore, a technical analysis could not be performed.Good faith effort attempts were made to try and retrieve the device; however, device was discarded.It was confirmed the device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.A risk review was completed and confirmed that the event of difficult to charge and premature discharge of battery can be related to ineffective or loss of therapy - short term was defined in the product risk documentation.This event type has been accounted for during analysis to support acceptable risk benefit for the product.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
 
Additional Manufacturer Narrative
Block d2b: additional pro code: qon block g4: additional premarket / 510 (k): p190006.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Axonics
Common Device Name
Stimulator, Electrical, Implantable, For Incontinence
Manufacturer (Section D)
Axonics Modulation Technologies, Inc.
15515 sand canyon ave
irvine CA 92618
Manufacturer (Section G)
Axonics Inc
15515 sand canyon ave
irvine CA 92618
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key50099807
Report Number2124215-2026-41234
Device Sequence Number13240069
Product Code EZW
UDI-Device Identifier10810005340066
UDI-Public(01)10810005340066(11)201223(17)231223
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P180046
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/23/2023
Device Model Number1101
Device Catalogue Number1101
Device Lot NumberAX1H168016
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/07/2026
Supplement Date Received by Manufacturer08/05/2026
Initial Report FDA Received Date07/31/2026
Supplement Report FDA Received Date08/31/2026
Date Device Manufactured12/23/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age37 YR
Patient SexFemale
-
-