It was reported that the log files were submitted to evaluate power spikes.It appeared there were no unusual events recorded in the log file event history.The mechanical circulatory support (mcs) equipment was operating as intended.The pump parameters trending showed power and flow trending upward with pulsatility index (pi) trending downward slightly.Per technical services, this trending could be suspect of pump thrombus.On (b)(6) 2026, the patient was admitted for elevated flows and powers (>8).Computed tomography (ct) showed narrowing of outflow graft.Upon inspection in the cardiac catheterization laboratory, narrowing and gradient were not substantial enough for team to stent the graft.During the procedure, flows and powers returned to normal, with flows in the 4 liter per minute range (lpm) and powers in 6's.Additional log files sent for review revealed that the power went down from 8.6 watts (w) to 6.1w and estimated flow from 8.6lpm to 4.5lpm on 14 jul 2026 at 15:13.Otherwise, there were no other notable alarm conditions or parameter changes in the log file.The ventricular assist device (vad) was functioning as intended.
|