The device was not returned for product analysis; however, the device was serviced at the facility of the customer by a field service engineer (fse).The fse found that the customer was using very strange scissors to cut fibers.The fse instructed the customer to order a fiber cutter, stripper, and a blast shield for the customer.The reported allegations of the console burning a lot of blast shields and fibers, and the laser not feeling powerful anymore was most likely determined to be related to incorrect fiber and blast shield handling.A review of the device history record/service history was performed.The record indicates that product family p120 serial number 50 installation and functional verification were performed on 12/21/2020, system last service performed was on 6/22/2026, the console remained fully functional, with no issues reported until the last service and/or complaint date.Based on service records, the issue is more likely due to use-related wear or field conditions rather than manufacturing or design at release.A risk review was completed and confirmed that the event of device - low power was defined in the risk documentation and is documented accordingly in the product record review.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.A review of the device instructions for use (ifu) was completed.The ifu includes specific states regarding the turning on the laser: additional accessories - fiber cutting scissors, optical fiber strippers, cleaving tool.Fiber burn back - fiber burn back may occur during prolonged procedures, especially when using higher power.1.Renew the fiber tip by stripping and cleaving the fiber.The ifu has no issues with translation, wording, or graphics and therefore revision is not required.Based on the information provided, all compiled information on this investigation determines that the most probable cause is unintended use error caused or contributed to event since the interaction between the user and device, or sample, caused or contributed to the error.This includes unintended inappropriate use of the device and incorrect sample preparation.
|