Model Number CDS0801-NTW
Medical Device Problem Code Incomplete Coaptation (2507)
Health Effect - Clinical Codes Dyspnea (1816); Mitral Valve Insufficiency/ Regurgitation (4451)
Date of Event 07/10/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported on (b)(6) 2026 that the patient presented with grade 4 degenerative mitral regurgitation (mr) for a mitraclip procedure.During the procedure, 2 mitraclips were successfully implanted.The mr was reduced to grade 2 post procedure.On (b)(6) 2026, follow up revealed that one clip had single leaflet device attachment (slda) from the posterior leaflet.Recurrent mr of grade 3-4+ was reported along with dyspnea.There was no clinically significant delay reported.
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to the reported event.Additionally, a review of the complaint history identified no similar complaints from the lot.Based on available information, a cause for the reported incomplete coaptation- single leaflet device attachment (slda) could not be determined.The reported recurrent mitral valve insufficiency/ regurgitation (mr) and dyspnea appear to be cascading effects of the reported slda.Mr and dyspnea are known possible complications associated with mitraclip procedures.The reported serious injury/ illness/ impairment was a result of case-specific circumstances.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MitraClip G5 Delivery System
Common Device Name
mitral valve repair devices
Manufacturer (Section D)
Abbott Medical
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
Abbott Medical Northpoint Reg 3020950818
1820 bastian court
westfield IN 46074
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
+1(651)756-5400
MDR Report Key50100356
Report Number2135147-2026-05236
Device Sequence Number9803947
Product Code NKM
UDI-Device Identifier05415067050748
UDI-Public(01)05415067050748(17)270726(10)60126A1074
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
PMA/510(K) Number
P100009
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/31/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCDS0801-NTW
Device Catalogue NumberCDS0801-NTW
Device Lot Number60126A1074
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/10/2026
Initial Report FDA Received Date07/31/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/26/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-