• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Establishment Labs S.A. MOTIVA IMPLANTS; Ergonomix Round SmoothSilk

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Establishment Labs S.A. MOTIVA IMPLANTS; Ergonomix Round SmoothSilk Back to Search Results
Catalog Number ERSM-160
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Impaired Healing (2378)
Type of Reportable Event Serious Injury
Event or Problem Description
Usa.Allegedly, a patient presents an exposed implant, that required an explant surgery.The investigation is ongoing.
 
Additional Manufacturer Narrative
As stated in the literature: "extrusion represents potential complications associated with the use of breast implants, it involves the implant coming through the skin, normally through the incision where it¿s been placed.Extrusion can happen for several reasons, such as the skin over the implant thinned too much leaving too little coverage for the implant, the implant used is too large, the incision did not heal well and broke open.Implant exposure due to poor tissue coverage frequently obligates the surgeon to remove the breast implant and begin anew (gargano, ciminello & de santis, 2009)".Per our directions for use, each patient must receive the establishment labs s.A.¿motiva implants®: information for the patient¿ during her surgical consultation, it is the surgeons´ responsibility to ensure that the patient completely understands the information regarding the risks, benefits, and recommendations associated with silicone gel-filled breast implant surgery, as well as the complications typical of any type of surgery.This document is available in motiva® website: ifu.Motiva.Health.The instructions for use in the directions for use were reviewed, to determinate if there are indications that ensure the prevention and good handling of the product.For this event, it is considered that the information is suitable for the section of surgical precautions as follows: extrusion: "breast implant extrusion, or breast implant exposure, occurs when the breast skin and tissues that are holding the implant fail, causing the implant to protrude through the skin and become exposed.It happens in less than 2% of patients; it is shortly after breast augmentation or down the road.Breast implant extrusion can occur for various reasons, such as improper wound healing due to an infection, trauma, too little soft-tissue coverage, oversized implant coupled with too little tissue coverage, or lack of blood supply.Breast-implant extrusion calls for surgery and removal of the implant".In addition, a review of the device history record was conducted, and it was concluded that there were no deviations in the manufacturing process of this device that would have caused or contributed to this incident".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTIVA IMPLANTS
Common Device Name
Ergonomix Round SmoothSilk
Manufacturer (Section D)
Establishment Labs S.A.
building b-25, coyol free zone
alajuela,alajuela 20113
CS  20113
Manufacturer (Section G)
Establishment Labs S.A.
building b-25, coyol free zone
alajuela,alajuela 20113
Manufacturer Contact
susana hidalgo
building b-25, coyol free zone
alajuela,alajuela 20113
+1(800)355-4164
MDR Report Key50100958
Report Number3012883202-2026-02712
Device Sequence Number17456160
Product Code FTR
UDI-Device Identifier07445161010308
UDI-Public(01)07445161010308(11)250221(17)300221
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberERSM-160
Device Lot Number25021707
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/10/2026
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date07/31/2026
Supplement Report FDA Received Date09/02/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/21/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-