Catalog Number 515105
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/10/2026
Type of Reportable Event Malfunction
Event or Problem Description
No additional information.
 
Additional Manufacturer Narrative
(b)(4) follow up mdr for device evaluation: it was reported that white particles were observed in the liquid.While the syringe involved in the event was not returned for evaluation, three photos, one video, one used protector and one unused protector were provided to our quality team for investigation.Upon review of the photos and video, a particle was observed within the syringe.The returned protectors were inspected, no damage or related defect was observed that could have contributed to this observation.A device history review was conducted for the reported protector lot, and no deviations or nonconformances related to the reported concern were identified during manufacturing.Fragmentation testing is performed during manufacturing to evaluate any particulates generated after ten activations.Testing results were reviewed for the reported lot and results were found to be acceptable.Testing was also performed on the returned samples, the product met required specification and no particles were generated during testing.Based on the photo and video evaluation, the composition of the observed particle could not be identified; therefore, its specific origin could not be determined at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Our quality team regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD Phaseal
Common Device Name
Closed antineoplastic and hazardous drug reconstitution and transfer system
Manufacturer (Section D)
BECTON DICKINSON, S.A.
camino de valdeolivia
s/n
san agustin de guadalix
Manufacturer (Section G)
BECTON DICKINSON, S.A.
camino de valdeolivia
s/n
san agustin de guadalix
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
+1(847)393-5694
MDR Report Key50101617
Report Number3003152976-2026-00345
Device Sequence Number18642942
Product Code ONB
UDI-Device Identifier00382905151057
UDI-Public(01)00382905151057(11)260310(17)310228(10)2603115
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K123213
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number515105
Device Lot Number2603115
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/03/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/09/2026
Supplement Date Received by Manufacturer09/30/2026
Initial Report FDA Received Date07/31/2026
Supplement Report FDA Received Date10/01/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/10/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-