Atrium Medical Corporation ADVANTA SLIDER GDS; Prosthesis, Vascular Graft, Of 6mm And Greater Diameter Back to Search Results
Model Number 22187
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Codes Pain (1994); Thrombosis/Thrombus (4440); Limb Pain (4836)
Date of Event 12/11/2022
Type of Reportable Event Serious Injury
Event or Problem Description
N/a.
 
Event or Problem Description
As part of the advanta vxt and flixene pmcf registry, a report was received related to a vxt graft.It was reported that the patient implanted with advanta vxt graft had developed pain in the lower limb that was present for about ten days and had worsened.The patient presented with a painful foot, sensory and motor deficits extending to the mid-shin, and a swollen and painful calf.On examination, the bypass graft was thrombosed.A ct scan demonstrated occlusion of the femoropopliteal bypass graft.A thrombectomy was performed and the adverse event resolved.
 
Additional Manufacturer Narrative
Additional information: h6.Based on the complaint investigation, a retrospective review was conducted as part of the post-market clinical follow-up (pmcf) registry for advanta vxt and flixene vascular grafts.The pmcf registry is a multicenter observational study evaluating the long-term safety and performance of grafts implanted between (b)(6) 2010, and (b)(6) 2021.During retrospective review of patient records, adverse events including thrombosis, graft occlusion, infection, restricted blood flow, and/or the need for additional surgical intervention were reported.The manufacturer performed an investigation in accordance with established complaint handling procedures.Review of the available information determined that the reported events are consistent with known complications associated with vascular graft implantation and are recognized adverse events identified in the device instructions for use (ifu aw011456).Review of complaints received through the pmcf registry demonstrated a common pattern in which reported events were associated with patient comorbidities, progression of underlying vascular disease, and anticipated procedural complications.No physical product, photographs, or other objective evidence were provided for evaluation, and there was no information indicating a device malfunction, manufacturing deficiency, or design-related issue.A device history record review was not required because no device malfunction or product nonconformance was alleged or identified.Complaint trending and risk management records were reviewed.The reported hazard-harm combinations are documented and adequately addressed within the dfmeca for eptfe vascular grafts.Although an increase in thrombosis and occlusion complaints was observed during pmcf data collection, the increase was attributed to the influx of retrospective cases identified through the registry rather than a change in device performance.No new or increased risk was identified, and no capa, field action, or other corrective action was initiated as a result of this investigation.Based on the available information, the manufacturer was unable to confirm a device-related cause for the reported event(s).The investigation concluded that the event was most likely associated with patient factors, disease progression, and/or known procedural risks.The device was determined to have operated within specification, and no evidence was identified to suggest that the advanta vxt or flixene vascular graft contributed to the event through a product deficiency.The investigation is complete and no further action is warranted at this time.
 
Additional Manufacturer Narrative
Due to character restrictions in block e1: event site name: (b)(6).A supplemental report will be submitted upon completion of our investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANTA SLIDER GDS
Common Device Name
Prosthesis, Vascular Graft, Of 6mm And Greater Diameter
Manufacturer (Section D)
Atrium Medical Corporation
40 continental blvd
merrimack NH
Manufacturer (Section G)
Atrium Medical Corporation
40 continental blvd
merrimack NH
Manufacturer Contact
lori gosselin
40 continental blvd
merrimack, NH 
MDR Report Key50101708
Report Number3011175548-2026-000009
Device Sequence Number17469204
Product Code DSY
UDI-Device Identifier00650862221879
UDI-Public(01)00650862221879(17)221030
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K992960
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/30/2022
Device Model Number22187
Device Catalogue Number22187
Device Lot Number452140
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/10/2026
Supplement Date Received by Manufacturer09/22/2026
Initial Report FDA Received Date07/31/2026
Supplement Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/01/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
unknown.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-