• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Integra MicroFrance S.A.S. HOOK CEV2295A 3PK N5 FOR HOOK HANDLE; Microfrance Instruments

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Integra MicroFrance S.A.S. HOOK CEV2295A 3PK N5 FOR HOOK HANDLE; Microfrance Instruments Back to Search Results
Catalog Number CEV2295A
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/07/2026
Type of Reportable Event Malfunction
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOOK CEV2295A 3PK N5 FOR HOOK HANDLE
Common Device Name
Microfrance Instruments
Manufacturer (Section D)
Integra MicroFrance S.A.S.
le pavillon
le pavillon
saint aubin le monial 03160
FR  03160
Manufacturer (Section G)
Integra MicroFrance S.A.S.
le pavillon
saint aubin le monial 03160
FR   03160
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 08540
+1(609)936-2319
MDR Report Key50101857
Report Number3003249645-2026-00029
Device Sequence Number10003383
Product Code GEI
UDI-Device Identifier10381780092568
UDI-Public10381780092568
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K210942
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Type of Report Initial,Followup
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberCEV2295A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/07/2026
Supplement Date Received by Manufacturer09/02/2026
Initial Report FDA Received Date07/31/2026
Supplement Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-