| Model Number |
3789 |
| Medical Device Problem Codes |
Improper or Incorrect Procedure or Method (2017); Wireless Communication Problem (3283)
|
| Health Effect - Clinical Code |
Inadequate Pain Relief (2388)
|
| Date of Event |
07/08/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Additional information indicates the patient stopped charging their ipg.Surgical intervention was undertaken on (b)(6) 2026 to address the issue wherein the ipg was explanted and replaced.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's ipg became unable to communicate with external devices.As a result, surgical intervention may be undertaken in the future.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated patient weight in section a4 - weight is estimated.Added explant date to section db6.The directions for use states that the implant must be charged every 30 days to avoid battery depletion. based on the information received, the cause of the reported incident is related to user error.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event is estimated.
|
| |
|