It was reported that a patient experienced a st segment elevation, ventricular tachycardia and an arrhythmia.During a pfa ablation procedure, a farawave catheter was selected, and pulmonary vein isolation (pvi) and posterior wall isolation (pwi) were completed in the left atrium.The patient was believed to be in typical atrial flutter at the time, which did not terminate with ablation.The procedure then moved to the right atrium, where a cavotricuspid isthmus (cti) line was created using the farawave catheter.The farawave catheter had been removed and reinserted twice during the case.No prophylactic nitroglycerin was administered, and no acute st segment changes were initially noted.While pacing to assess cti block, the patient developed ventricular tachycardia (vt).Initially, it was suspected that pacing induced the vt, however, review of the intracardiac electrograms suggested pacing was not the cause.Review of the ecgs demonstrated st segment elevation in leads v1 and v2 that was already present on the pre procedure ecg, indicating the finding was likely preexisting rather than procedure related.The vt morphology was consistent with an origin from the left ventricular apex.The patient had a history of atrial fibrillation/atrial flutter, which was the indication for ablation; however, vt was not a pre-existing arrhythmia and was not the target of the procedure.To treat the arrhythmia, the patient was successfully cardioverted out of vt and subsequently administered nitroglycerin.In the physician opinion, it remains unknown whether the device or procedure caused or contributed to the event.The patient experienced complications of ventricular arrhythmia and st segment elevation, although the st segment elevation appeared to be present at baseline.Due to the event, the patient was admitted overnight beyond the standard of care, and a cardiac magnetic resonance imaging was ordered for further evaluation.The patient had not yet been discharged at the time of reporting and was expected to make a full recovery.
|