It was reported on 07 july 2026 that a 30-25mm amplatzer talisman pfo occluder was chosen for implant using a 9f amplatzer talisman delivery system.The defect was reported to be large patent formen ovale (pfo) with a width of approximately 8mm and a septum primum aneurysm.Transesophageal echocardiogram (tee) was used during the procedure.A stability check was reported to have been performed, with no leak detected around the device.No difficulty with implantation, including multiple recaptures or repositioning, was reported, and no rhythm change occurred during the procedure.The device was successfully deployed and released.A few minutes after release, the device embolized and migrated to the ascending aorta.A transthoracic echocardiogram (tte) reportedly identified the embolization.No partial or complete obstruction of blood flow was reported, and the patient reportedly experienced no clinical signs or symptoms associated with the event.It was reported that a percutaneous retrieval attempt was unsuccessful and was followed by surgical retrieval via vascular surgery.
|