Upon receipt at our quality assurance laboratory, the cuff of this artificial urinary sphincter (aus) underwent a thorough analysis.Visual examination revealed that the cuff shell was worn from wear at a fold on both sides of the median of the cuff.A hole from avulsion was found along the lateral edge in the median of the cuff.The cuff had a leak due to a hole from avulsion.The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications.A review of the device instructions for use (ifu) was completed and did not reveal any evidence of device misuse, off label use, or failure to follow instructions.A risk review was completed and confirmed that the event of fluid leak was defined in the product risk documentation.This event type has been accounted for during analysis to support acceptable risk benefit for the product.Based on the information available and analysis results, a conclusion code of cause traced to component failure was assigned to this investigation.
|