| Model Number |
EEA25 |
| Medical Device Problem Codes |
Failure to Form Staple (2579); Human-Device Interface Problem (2949)
|
| Health Effect - Clinical Codes |
Failure to Anastomose (1028); Laceration(s) (1946)
|
| Date of Event |
07/08/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during an esophagectomy, a single use stapler was difficult or unable to close during set-up as the white safety tab was stuck.It was also reported that, during firing, there was an incomplete staple line with partial firing.It was reported that the issue was addressed by hand sewing and converting the procedure from minimally invasive surgery to an open approach.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|