| Model Number |
251 (+ 22 D) |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/02/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Event occurred in china.Damaged haptic after/during implantation.The patient underwent cataract surgery with implantation of an intraocular lens (iol).After the iol was injected and implanted during the procedure, it was noted that the distal end of the posterior haptic had fractured.The broken haptic fragment remained inside the injector, resulting in loss of integrity of the implanted iol's posterior haptic.Subsequent examination revealed no retained intraocular foreign body and no evidence of ocular structural damage at present.However, there remains a potential risk of iol decentration/dislocation and capsular bag instability.Problem code: a0414 - material split, cut or torn.
|
| |
|
Additional Manufacturer Narrative
|
|
This initial emdr is being submitted to the fda for a reportable event that occurred outside the usa.Haptic damage is indicated as a potential malfunction related to the iol, as stated under the warnings section of the product's instructions for use (ifu).Regarding section h6 - manufacturer's codes for: type of investigation, findings, and conclusion are pending completion of product investigation.Once the product investigation is completed, a follow-up report will be submitted to fda which will include the manufacturer's codes for type of investigation, findings, and conclusion.
|
| |
|