It was reported to boston scientific corporation that a wallflex esophageal stent was to be implanted in the esophagus during an esophageal stent placement procedure performed on (b)(6)2026.During the preparation prior to the procedure, the stent was opened and a white string was observed exiting the distal cap directly out of the box.The device was not used, and another wallflex esophageal stent was used to complete the procedure.There were no patient complications as a result of this event.
|
|
Block h6:
imdrf device code a180104 captures the reportable event of device - foreign material present in device.Block h11:
investigation results:
boston scientific corporation confirmed the reported event of device foreign material present in device.The investigation concluded that there is not enough evidence to determine whether the reported event device foreign material present in device was related to a manufacturing issue, packaging condition, handling, storage, transportation, or another contributing factor.Additionally, the evidence from the product record review did not identify a potential product quality problem or new patient harm.Without the ability to examine the device, the most probable cause of the event remains unknown.Device history record review:
it was confirmed this device met manufacturing specification prior to distribution and there were no manufacturing deviations which have contributed to the reported event
device technical analysis:
the device was not returned; therefore, product analysis could not be performed.However, a media inspection was performed on a photograph provided by the complainant.The photograph showed evidence of the distal tip of the device with a foreign material protruding from the tip area.Similar complaint review:
a similar complaint review was completed.There were no emerging trends identified that warrant further action.Investigation conclusion:
based on the available information, there is not enough evidence to confirm the cause of reported event.Without proper evaluation of the complaint device, cause not established is selected as the investigation conclusion code.
|