During procedure for a pulse field ablation (pfa) and a watchman left atrial appendage closure (laac) flx pro procedure to treat atrial fibrillation a versacross connect access solution for faradrive and a farawave ablation catheter were selected for use.Procedure started at 13:00, with a patient cleared for pericardial effusion at the beginning of case by the physician.Physician experienced difficulty getting the versacross into the superior vena cava (svc) due to pacer leads, wire was exchange for j-tipped (non-bsc) wire to navigate through leads, once inside svc wire was exchanged again.Upon dropping down onto the fossa to perform transeptal puncture, with no felt resistance or difficulty, it was noted the septum was thickened, which lead to an unsuccessful first attempt to cross, second attempt was successful.Then, patient had ablation performed with no symptoms and/or decompensation.Watchman sheath was inserted at 13:58, pigtail was placed into laa and angio was performed.Drawing was marked on fluoroscopy and selected 31mm, it was prepared and inserted through trusteer sheath.With the use of tee watchman was deployed and ran through pass criteria.Implant was successful and device was released with no recaptures.Procedure was completed successfully with original devices.Final pericardial effusion check revealed large pericardial effusion collecting beneath the right ventricle, pericardiocentesis was performed.Patient was admitted the patient to the step-down cardiac unit with drain in place, requiring hospitalization beyond standard of care.Patient was discharged.The patient is expected to fully recover from the event.The physician stated that reported that during the first attempt to perform transeptal puncture rf wire crossed into the pericardial space and patient developed the pericardial effusion since that point in the procedure.The device is not expected to return as it was disposed.
|