| Model Number |
GZ-130P |
| Medical Device Problem Codes |
Application Program Problem (2880); Unintended Application Program Shut Down (4032)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/08/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The biomedical engineer (bme) reported that the gz telemetry transmitter is constantly rebooting and not staying on.The bme swapped out the batteries with a new set and the issue still persists.Not in patient use.
|
| |
|
Additional Manufacturer Narrative
|
|
The biomedical engineer (bme) reported that the gz telemetry transmitter is constantly rebooting and not staying on.The bme swapped out the batteries with a new set and the issue still persists.Not in patient use.Investigation conclusion:
the device was returned to nihon kohden for evaluation.The evaluation remarked on signs of liquid exposure and poor overall condition.The reported issue was replicated, and the cause was attributed to circuit board failure.The device was deemed ineligible for repair and designated for scrapping.Additional information:
b4 date of this report
d9 device available for evaluation
g3 date received by manufacturer
g6 type of report
h2 if follow-up, what type?
h3 device evaluated by manufacturer
h6 event problem and evaluation codes.H11 additional manufacturer narrative.
|
| |
|