It was reported that after an initial bunion surgery on (b)(6) 2025, all hardware was removed on (b)(6) 2026 due to non-union and a broken staple.No other patient outcomes were reported as a result of this event.Additional information indicated that patient was overweight with other comorbidities and was fairly active.The device history records were reviewed, and no issues were identified during the manufacture and release of the devices that could have contributed to what was reported.A number of factors could have contributed to what was reported and although the most likely cause cannot be determined based on the available information, the broken plate was likely subjected to excessive bending stresses which resulted in its breakage due to the patient's weight, comorbidities, and level of activity.The company will supplement the mdr as necessary and appropriate.
|