• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Datascope Corp. CS300 Intra-Aortic Balloon Pump, English, 220V; System, Balloon, Intra-Aortic And Control

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Datascope Corp. CS300 Intra-Aortic Balloon Pump, English, 220V; System, Balloon, Intra-Aortic And Control Back to Search Results
Model Number 0998-00-3023-55
Medical Device Problem Code Battery Problem (2885)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/13/2026
Type of Reportable Event Malfunction
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 Intra-Aortic Balloon Pump, English, 220V
Common Device Name
System, Balloon, Intra-Aortic And Control
Manufacturer (Section D)
Datascope Corp.
1300 macarthur blvd
mahwah NJ 07430
Manufacturer (Section G)
Datascope Corp.
1300 macarthur blvd
mahwah NJ 07430
Manufacturer Contact
ewelina chorlowska
1300 macarthur blvd
mahwah, NJ 07430
MDR Report Key50113666
Report Number2249723-2026-0003697
Device Sequence Number10761192
Product Code DSP
UDI-Device Identifier10607567112541
UDI-Public(01)10607567112541
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K063525
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/12/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0998-00-3023-55
Device Catalogue Number0998-00-3023-55
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/13/2026
Supplement Date Received by Manufacturer09/11/2026
09/12/2026
Initial Report FDA Received Date08/03/2026
Supplement Report FDA Received Date09/12/2026
09/12/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/11/2018
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-