| Model Number |
02.07.0510FUC |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Joint Laxity (4526)
|
| Date of Event |
07/22/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The patient had knee instability and the cause is unknown.At about 14 years and 10 months post primary the surgeon upsized the insert and the surgery was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
Batch review performed on 27 july 2026: lot 110360: 25 items manufactured and released on 02-may-2011.Expiration date: 31-mar-2016.No anomalies found related to the problem.To date, all items of the same lot have been sold with no similar reported event during the period of review.Root cause:the surgeon revised liner height, which is a common practice to restore the joint stability.There is no indication that any potential issue with the device may have caused or contributed to the event.
|
| |
|