• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medacta International SA GMK Primary Total Knee System; Patella resurfacing S.2 (New generation)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medacta International SA GMK Primary Total Knee System; Patella resurfacing S.2 (New generation) Back to Search Results
Model Number 02.07.0034RP
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 07/21/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery due to patella maltracking and dislocation after about 1 year and 4 months from primary.All knee implants were revised using competitors implants.The primary was performed using myknee patient match cutting block.
 
Additional Manufacturer Narrative
Patient match planning review knee the analysis of the process found out no errors: each step of the mysolutions process has been performed accordingly with the working instructions batch review performed on 27 july 2026 lot 2425998: (b)(6) items manufactured and released on 31-oct-2024.Expiration date: 2029-oct-15.No anomalies found related to the problem.To date, all items of the same lot have been sold with no similar reported event during the period of review.Root cause: no definitive root cause could be established.The myknee planning review identified no errors or potentially related manufacturing issues; however, the absence of post-operative imaging prevented assessment of the actual implant positioning and other clinical or surgical factors.There is no indication that any potential issue with the device may have caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GMK Primary Total Knee System
Common Device Name
Patella resurfacing S.2 (New generation)
Manufacturer (Section D)
Medacta International SA
strada regina
castel san pietro 6874
SZ  6874
Manufacturer Contact
marco giannessi
strada regina
castel san pietro 6874
SZ   6874
MDR Report Key50114124
Report Number3005180920-2026-00806
Device Sequence Number5468342
Product Code JWH
UDI-Device Identifier07630030815751
UDI-Public07630030815751
Combination Product (Y/N)N
PMA/510(K) Number
K090988
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 08/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number02.07.0034RP
Device Lot Number2425998
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/21/2026
Initial Report FDA Received Date08/03/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/31/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Patient SexMale
-
-