| Model Number |
01.18.152 |
| Medical Device Problem Code |
Implant subsidence (4078)
|
| Health Effect - Clinical Code |
Implant Pain (4561)
|
| Date of Event |
07/21/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Revision surgery performed at 20 days from primary due to femoral stem subsidence.Stem, head and liner successfully revised.
|
| |
|
Additional Manufacturer Narrative
|
|
Batch review performed on 22 jul 2026 lot 2435536: (b)(4) items manufactured and released on 04/04/2025.Expiration date: 12/03/2030.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Root cause:implant subsidence is a possible literature-described adverse event after primary joint arthroplasties and causes are often unknown.The event appears multifactorial and may be related to loss or lack of stem fixation/osseointegration, patient or surgical factors.There is no indication that any potential issue with the device may have caused or contributed to the event.
|
| |
|