| Model Number |
02.18.EIF4.08.RM |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Medical device site infection (4845)
|
| Date of Event |
07/17/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Revision surgery due to infection after about 1 month from primary.The surgeon performed a washout and revised the poly.
|
| |
|
Additional Manufacturer Narrative
|
|
Batch review performed on 03 august 2026 02.18.Eif4.08.Rm moto medial e-cross tibial insert s4 rm - h8 (k213071) lot.2517155: (b)(4) items manufactured and released on 16 october 2025.Expiration date: 29 september 2030.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Root cause:infection is a known possible complication of any surgery.Although no root cause can be established, there is no indication that any issue with the device may have caused or contributed to the event, and the document review does not indicate any potential manufacturing related cause.
|
| |
|