Revision at 5 days from primary surgery for bone fracture.The reason of fracture is unknown.The shell was found in a different position from the primary, likely as a secondary effect of the fracture.Shell, liner, and head revised successfully.
|
|
Batch review performed on 24 july 2026: lot 2516641: (b)(4) items manufactured and released on 03-dec-2025.Expiration date: 23-nov-2030.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Root cause:based on the available information, the reported bone fracture is considered the most likely cause of the change in shell position and the subsequent early revision.However, in the absence of further clinical or radiographic information, the relationship between the fracture and shell displacement cannot be conclusively established.The batch review did not identify any potentially related manufacturing issue.
|