Model Number 02.07.1202L
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Implant Pain (4561); Aseptic loosening (4883)
Date of Event 07/21/2026
Type of Reportable Event Serious Injury
Event or Problem Description
At about 3 years and 7 months from the primary, the patient came in reporing paindue to loosening of the tibial trayand the cause is unknown.The surgeon revised all components and the surgery was completed successfully.
 
Additional Manufacturer Narrative
Batch review performed on 28 july 2026 gmk-sphere 02.07.1202l tibial tray fix cemented s.2l (k090988) lot 2008102: 20 items manufactured and released on 10-dec-2020.Expiration date: 29-nov-2025.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Root cause:aseptic loosening is possible literature described adverse event after primary joint arthroplasties and causes are often unknown.There is no indication that any potential issue with the device may have caused or contributed to the event and the investigation does not indicate any potential manufacturing related issue or systemic deficiency.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GMK Primary Total Knee System
Common Device Name
Tibial tray fix cemented S.2L
Manufacturer (Section D)
Medacta International SA
strada regina
castel san pietro 6874
SZ  6874
Manufacturer Contact
marco giannessi
strada regina
castel san pietro 6874
SZ   6874
MDR Report Key50114139
Report Number3005180920-2026-00830
Device Sequence Number14778654
Product Code JWH
UDI-Device Identifier07630030819865
UDI-Public07630030819865
Combination Product (Y/N)N
PMA/510(K) Number
K090988
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 08/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/29/2025
Device Model Number02.07.1202L
Device Lot Number2008102
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/21/2026
Initial Report FDA Received Date08/03/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/10/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-