BOSTON SCIENTIFIC CORPORATION WATCHMAN FLX? Pro; system, appendage closure, left atrial
|
Back to Search Results |
|
| Model Number |
M635WU60350 |
| Medical Device Problem Code |
Failure to Seal (4070)
|
| Health Effect - Clinical Code |
Stroke/CVA (1770)
|
| Date of Event |
05/15/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that cerebral vascular accident occurred.A left atrial appendage (laa) closure procedure was performed, and a 35mm watchman flx pro closure device was implanted on (b)(6) 2025.The patient was discharged.It was reported that the patient experienced a stroke within the two months preceding (b)(6) 2026, the reporter was unable to provide the exact date of event.Following the closure device implant, the patient received a post 45-day computed tomography (ct) scan on (b)(6) 2025 that showed a 4mm peri device leak.Subsequently, a transesophageal echocardiogram (tee) performed on (b)(6) 2026, revealed a 3.5mm gap on the superior access, a small lobe uncovered, and a 2mm posterior device leak.The reporter was unable to provide information regarding the post-procedure medication regimen initiated after the december ct scan or whether any interventions were performed in response to the ct findings.The reporter was also unaware of any medication changes or interventions following the june tee findings.
|
| |
|
Event or Problem Description
|
|
It was reported that cerebral vascular accident occurred.A left atrial appendage (laa) closure procedure was performed, and a 35mm watchman flx pro closure device was implanted on (b)(6)2025-.The patient was discharged.It was reported that the patient experienced a stroke within the two months preceding (b)(6)2026, the reporter was unable to provide the exact date of event.Following the closure device implant, the patient received a post 45-day computed tomography (ct) scan on (b)(6)2025 that showed a 4mm peri device leak.Subsequently, a transesophageal echocardiogram (tee) performed on (b)(6)2026, revealed a 3.5mm gap on the superior access, a small lobe uncovered, and a 2mm posterior device leak.The reporter was unable to provide information regarding the post-procedure medication regimen initiated after the december ct scan or whether any interventions were performed in response to the ct findings.The reporter was also unaware of any medication changes or interventions following the june tee findings.
|
| |
|
Additional Manufacturer Narrative
|
|
B3 event date is an approximate date as the actual event date is not known.With all the available information, boston scientific (bsc) concludes:
device technical analysis
the watchman flx? pro remains implanted; therefore, returned device analysis could not be completed.Device history record review
a review was completed of the device history records (dhr) for the watchman flx? pro, part # m635wu60350 batch # 0034387693.The dhr review confirmed that the device met all material, assembly, and performance specifications prior to distribution.Labeling review
there was no evidence to suggest that the device was used in a manner inconsistent with the labeling.Watchman flx? pro instructions for use (ifu) lists potential complications, risks and adverse events that may be associated with the use of this device which include improper seal of the appendage, stroke (cva).Risk review
a risk review was completed and confirmed that the events of improper seal of the appendage and cerebral vascular accident (cva) were defined in the risk documentation.These event types have been accounted for during product risk analysis to support acceptable risk benefit for the product.Investigation conclusion
based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
|
| |
|
Additional Manufacturer Narrative
|
|
B3 event date is an approximate date as the actual event date is not known.
|
| |
|