| Model Number |
103606-620 |
| Medical Device Problem Code |
Imprecision (1307)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/07/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On august 13th 2025, senseonics was made aware of an incident where user experienced sensor inaccuracy.No injury or medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This mdr is submitted retrospectively following an internal review and updated best practices in the reporting criteria.The user reported differences between sg and bg readings.Escalation analysis indicated that the system was working within expectations.The observed discrepancies were attributable to lag between the sg and bg inherent to sensing technology.Customer care advised the user to continue using the system and to reach out if they had any further concerns.
|
| |
|