Cayenne Medical QUATTRO X3 ANCHOR 5.5MM; Fastener, Fixation, Nondegradable, Soft Tissue Back to Search Results
Catalog Number CM-9255X3
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/08/2026
Type of Reportable Event Malfunction
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected.Updated: d9; h2; h3; h6.Reported event was able to be confirmed due to device returned.It was confirmed that the suture thread was fractured and not the anchor.Visual examination of the returned device identified that the suture was returned with a knot near end of the suture where the end is frayed.Confirmed that there are 3 bundles of multifilament tracer included in the suture.Measured the suture length with a ruler which measures under 19".Also measured the diameter of the suture with micrometers and the measurement was within the specs listed in the attached raw material spec.No additional part of the device was returned product was returned or additional pictures provided; further visual and dimensional evaluations could not be performed.Medical records were not provided.Device history records were reviewed and no discrepancies related to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUATTRO X3 ANCHOR 5.5MM
Common Device Name
Fastener, Fixation, Nondegradable, Soft Tissue
Manufacturer (Section D)
Cayenne Medical
16597 n 92nd street
scottsdale AZ 85260
Manufacturer (Section G)
Cayenne Medical
16597 n 92nd street
scottsdale AZ 85260
Manufacturer Contact
jennifer rapsavage
56 e. bell dr
warsaw, IN 46582
+1(574)526-0384
MDR Report Key50115038
Report Number3006108336-2026-00019
Device Sequence Number20160184
Product Code MBI
UDI-Device Identifier00887868270284
UDI-Public(01)00887868270284(17)310324(10)67806008
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K112814
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberCM-9255X3
Device Lot Number67806008
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/23/2026
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/08/2026
Supplement Date Received by Manufacturer09/25/2026
Initial Report FDA Received Date08/03/2026
Supplement Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/24/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-