| Model Number |
72404232-10 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/26/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this inflatable penile prosthesis (ipp) was no longer working.Upon inspection, a fracture to the cylinder was noticed.As a result, the device was explanted.No patient complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this inflatable penile prosthesis (ipp) was no longer working.Upon inspection, a fracture to the cylinder was noticed.As a result, the device was explanted.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is not available for analysis; therefore, no physical analysis of the product could be performed.The reported device performance allegation cannot be confirmed.The device history record (dhr) review was unable to be performed as the lot number was not provided.A review of the device instructions for use (ifu) was completed and did not reveal any evidence of device misuse, off label use, or failure to follow instructions.A risk review was completed and confirmed that the event of cylinder hole/perforate and mechanical issue was defined in the product risk documentation.This event type has been accounted for during analysis to support acceptable risk benefit for the product.Based on the information available the cause that contributed to the reported event cannot be established.Therefore, a conclusion code of cause not established was assigned to this investigation.Detailed product information was not provided to bsc.We completed a good faith effort to obtain the information.Because the product is unknown at this time, we are unable to provide the complete unique identifier (udi) # and other product specific information.If additional details become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Detailed product information was not provided to bsc.We completed a good faith effort to obtain the information.Because the product is unknown at this time, we are unable to provide the complete unique identifier (udi) # and other product specific information.If additional details become available, a supplemental report will be submitted.
|
| |
|
Search Alerts/Recalls
|