| Catalog Number |
2000342 |
| Medical Device Problem Codes |
Failure to Advance (2524); Deformation Due to Compressive Stress (2889); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/22/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Clinical narrative:
impella rp flex was inserted percutaneously via right internal jugular vein access site in an 59-year-old female patient presenting with post-operative cardiothoracic surgery.The patient¿s comorbidities were prior coronary artery bypass graft (cabg) and known coronary artery disease (cad).The patient¿s clinical state prior to initiation of support was society for cardiovascular angiography and interventions (scai) shock stage e.The patient was medically supported by inotropes, vasopressors and respiratory support.During vascular access, difficulty was encountered while advancing the 23 fr peel-away introducer sheath.The physician removed the sheath and observed that the sheath tip was buckled inward/collapsed.A replacement introducer from a second rp kit was obtained and inserted successfully without further issue.Following rp flex insertion, a ¿placement signal not reliable¿ alarm was observed on the automated impella controller (aic) resulting in no placement signal or flows displayed.The lack of flow and placement signals being displayed had no noted impact on hemodynamic support provided or the patient.Purge pressure and purge flow remained within appropriate ranges.
|
| |
|
Additional Manufacturer Narrative
|
|
The pump and introducer will be returned for investigation.This is one of two related reports.This report represents the introducer and another report represents the impella rp flex.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by abiomed inc, or its employees that the report constitutes an admission that the product, abiomed inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|