It was reported that the coil stretched out of the target location.A 10mm x 50cm embold? fibered coil was selected for use in an embolization procedure of the non-tortuous left ovarian vein.The target vessel diameter was 6-7 mm.The patient was being treated for pelvic congestion.The distal left ovarian vein was embolized with two embold? fibered coils successfully.Afterwards, the proximal ovarian vein was embolized with two embold? fibered coils.It appeared normal and the whole delivery system and microcatheter were pulled out of the ovarian vein; however, in doing so, the second to last coil became entangled with either the microcatheter or the 5fr catheter, resulting in the coil unraveling with a strand stretched out.A fine strand of the coil extended all the way back into the sheath through the inferior vena cava (ivc).During removal of the microcatheter, tension on this strand pulled the proximal portion of the coil up into the renal vein.The team was able to reposition the proximal portion of the coil back into the ovarian vein; however, despite several hours of effort, using various guidewires and snares, they were unable to retrieve the remaining strand that extended out into the renal vein nor were they able to push it back into the ovarian vein.The coil never broke, but the stretched strand was cut down at the groin access.The surgeon decided to leave the remainder of the stretched coil in place due to its very small size.The procedure was prolonged as a result of this event.The patient was put on aspirin for the next three months with an expectation to recover fully.
|