| Model Number |
IO-300 |
| Medical Device Problem Codes |
Device Emits Odor (1425); Overheating of Device (1437)
|
| Health Effect - Clinical Code |
Unspecified Respiratory Problem (4464)
|
| Date of Event |
07/13/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient's device was overheating and a burning smell was noted.The patient stated they were hospitalized due to the event.A replacement device was sent.No additional information is available at this time as attempt to follow-up with the patient for more information was unsuccessful.
|
| |
|
Additional Manufacturer Narrative
|
|
B3: date of event is estimated.The inogen team attempted to contact the patient for further information, but the patient declined.There is no confirmation for the return reason of system hot.Zeolite is found contaminated which possibly caused when it absorbs the moisture.
|
| |
|