Visual inspection: during the investigation, no significant deformations or damage of the valves was determined.Permeability test: a permeability test has shown that the valve is permeable.Computer controlled test: to investigate the claim of over- drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valve is tested in both horizontal and vertical positions.The results show that the valve operates within the accepted tolerance in both positions.Adjustment test: the m.Blue was tested and is adjustable to all specified pressures.Braking force and brake function test: the brake functionality test has shown that the brake function is fully operational, and the braking force is within the given tolerances.Internal inspection: after dismantling of the valve, no deposits were found.Result: according to the results of investigation, no functional deviations or other abnormalities of the product was determined.The product operated within all specifications.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory action is required from our point of view.
|