• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Abbott Medical TriClip G5 Delivery System; tricuspid valve repair device, percutaneously delivered

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Medical TriClip G5 Delivery System; tricuspid valve repair device, percutaneously delivered Back to Search Results
Model Number TCDS0801-XT
Medical Device Problem Codes Difficult to Remove (1528); Unintended Movement (3026)
Health Effect - Clinical Code Unspecified Tissue Injury (4559)
Date of Event 07/09/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on (b)(6) 2026, a patient presented with grade 5 functional tricuspid regurgitation (tr) and rotated heart for a triclip procedure.During the procedure, triclip was advanced into the tricuspid valve and when tried to retract into the right atrium, clip got stuck in the chordae and flail segment was observed.Lock knob was removed, lock was tested and the clip was opened to 60 degrees.Physician unlocked the clip, relocked, closed and tested the lock, clip opened to 60 degrees.Physician unlocked the clip, opened to 120 degrees, relocked and closed the clip.Physician tested twice before deployment sequence and was implanted.All steps were performed as per ifu.The final tr was grade 2.There were no adverse patient effects or clinically significant delays reported.
 
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TriClip G5 Delivery System
Common Device Name
tricuspid valve repair device, percutaneously delivered
Manufacturer (Section D)
Abbott Medical
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, Reg # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
+1(651)756-5400
MDR Report Key50119852
Report Number2135147-2026-05279
Device Sequence Number10480992
Product Code NPS
UDI-Device Identifier05415067050847
UDI-Public(01)05415067050847(17)270518(10)51118R1019
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P230007
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTCDS0801-XT
Device Catalogue NumberTCDS0801-XT
Device Lot Number51118R1019
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/09/2026
Supplement Date Received by Manufacturer08/24/2026
Initial Report FDA Received Date08/03/2026
Supplement Report FDA Received Date08/28/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/18/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Triclip steerable guide catheter
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-