| Model Number |
479878 |
| Medical Device Problem Codes |
Extracardiac stimulation (1463); High Capture Threshold (3266)
|
| Health Effect - Clinical Code |
Device Overstimulation of Tissue (1991)
|
| Date of Event |
07/16/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the left ventricular (lv) lead exhibited slightly elevated capture thresholds since implant fifteen months prior.The patient experienced diaphragmatic stimulation in all lead vectors.The lead was explanted and replaced with a left bundle branch lead. no further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Continuation of d10: 457445 lead implanted (b)(6)2014.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|