| Model Number |
SGC0301 |
| Medical Device Problem Code |
Air/Gas in Device (4062)
|
| Health Effect - Clinical Codes |
Air Embolism (1697); Low Blood Pressure/ Hypotension (1914)
|
| Date of Event |
07/12/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that on 12 july 2026, a patient presented with grade 4+ functional mitral regurgitation (mr) and an enlarged atrium for a mitraclip procedure.During the procedure, when the steerable guide catheter (sgc) was advanced within the interatrial septum into the left atrium, air bubbles were visualized within the left atrium on echocardiography.The patient presented with st-segment elevation and an air embolism was believed to have occurred but was unable to be visualized.The sgc was removed from the patient and the patient stabilized.A replacement sgc was selected and advanced successfully and three mitraclips were implanted successfully.The procedure was discontinued, the final mr was reduced to grade 1+.There were no adverse patient effects or clinically significant delays reported.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026, a patient presented with grade 4+ functional mitral regurgitation (mr) and an enlarged atrium for a mitraclip procedure.During the procedure, when the steerable guide catheter (sgc) was advanced within the interatrial septum into the left atrium, air bubbles were visualized within the left atrium on echocardiography.The patient presented with st-segment elevation and an air embolism was believed to have occurred but was unable to be visualized.The sgc was removed from the patient and the patient stabilized.A replacement sgc was selected and advanced successfully and three mitraclips were implanted successfully.The procedure was discontinued, the final mr was reduced to grade 1+.There were no adverse patient effects or clinically significant delays reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.Based on available information and without the device to analyze, the cause of the reported air in the device was unable to be determined.The reported unexpected medical intervention was due to case-specific circumstances.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
| |
|